LGD Compact-A C2H2/H2O

Axetris’ Laser Gas Detection modules LGD Compact-A C2H2/H2O are a suitable solution for several measurement tasks in air quality and safety monitoring. With a continuous and real time measurement several processes in the chemical industries can be monitored and improved.

 

Benefits at a Glance:

- Designed for portable instruments
- Sub ppm detection limits
- Real-time measurement

Technical Data

Details C2H2 0-500 ppm H2O 0-10 vol 0.7 Hz + Add to Watchlist + Add to Comparison Print Specifications

Article No. *605.970*

Attribute Unit Value
Target Gas C2H2
Second Target Gas H2O
Measurement Type (EN) extractive, ambient / (DE) extraktiv, ambient
Measuring Range C2H2 ppm 0 - 500
Measuring Range Details C2H2 (EN) Calibration up to 100 ppm / (DE) Kalibrierung bis zu 100 ppm
Measuring Range H2O % vol. 0 - 10
Measuring Range Details H2O (EN) Calibration to full scale (FS) range / (DE) Kalibrierung auf den vollen Skalenbereich (FS)
Accuracy C2H2 ppm ± 1
Accuracy H2O % ± 0.15
Accuracy Details (EN) The accuracy is defined as the difference between the mean response during a 2 min time interval and the reference value at reference conditions. Variations of the operating temperature and pressure can affect the accuracy of the system. / (DE) Die Genauigkeit ist definiert als die Differenz zwischen dem mittleren Ansprechen während eines Zeitintervalls von 2 Minuten und dem Referenzwert bei Referenzbedingungen. Schwankungen der Betriebstemperatur und des Betriebsdrucks können die Genauigkeit des Systems beeinträchtigen.
Precision 2σ C2H2 ppm ≤ 0.8
Precision 2σ H2O % ≤ 0.1
Precision Details (EN) The precision is defined as 2 times the standard deviation s of a 2 min measurement at stable reference conditions over the full measuring range. The precision of the system can be improved by mean of averaging. / (DE) Die Präzision ist definiert als das Zweifache der Standardabweichung s einer 2-minütigen Messung unter stabilen Referenzbedingungen über den gesamten Messbereich. Die Genauigkeit des Systems kann durch Mittelwertbildung verbessert werden.
Sampling Rate Hz 0.7
Temperature Limitation °C +15 to +55