LGD F200-A NH3

Axetris’ Laser Gas Detection modules LGD F200 NH3 can be a benefit for many environmental applications like monitoring of NH3 emissions in agriculture and livestock and also for quality control  in biogas production. Furthermore, the detection of NH3 in human breath can help to detect diseases at an early stage and thereby contributes to cost savings in healthcare systems.

 

Benefits at a Glance:

-High selectivity to target gas
-sub ppm sensitivity
-Real-time measurement
 

Technical Data

Details NH3 0-100 ppm; 1 Hz + Add to Watchlist + Add to Comparison Print Specifications

Article No. *602.330*

Attribute Unit Value
Target Gas NH3
Measurement Type extractive, ambient
Measuring Range ppm 0 - 100
Measuring Range Details Calibration to full scale (FS) range, standard is 100 ppm, (possible up to 40'000 ppm, no warranty of precision)
Accuracy % +/-2
Accuracy Details Full scale reading depending on integration stability (temperature & pressure)
Precision 2σ ppm 0.4
Precision Details The precision is defined as 2 times the standard deviation sigma of a 5 minute measurement at stable reference conditions. System performance is precision-limited (not accuracy-limited) when the system temperature is stable.
Sampling Rate Hz 1

Scope of delivery

LGD F200 OEM-module, manual, accessories